YOKOSUKA YK09-08 Expendable Bathythermograph (XBT)

Last Modified: 2019-09-21
: YK09-08
: Processed (DMO)-QCed
: JAMSTEC
: Depth, Temperature
:
OCEANS > OCEAN TEMPERATURE > WATER TEMPERATURE

For Using Data

Data Management Office
See Terms and Conditions about constrain of use.
See Terms and Conditions about data citation.
Instrument:
XBT/XCTD
Using XBT (eXpendable Bathy Thermograph) system, the vertical distribution of water temperature is observed during free fall of its probe part in the seawater. On board, the analogue signal is converted to the temperature by data processor and the data is stored in PC. Depth data is calculated from the elapsed time.

(1) Launcher
Hand launcher
Manufacturer : Sippican,Inc.
Operation area : Rear upper deck

(2) Converter
Manufacturer : Tsurumi Seiki Co., LTD.
Location : Investigation room
Sampling rate : 50 msec

(3) XBT probe specifications
Probe Type TSK T-5 TSK T-6 TSK T-7 TSK T-10
Temperature range [deg-C] -2 to 35
Temperature accuracy [deg-C] +/- 0.2
Temperature resolution [deg-C] 0.01
Measurment depth [m] 1830 460 760 300
Depth accuracy [m] 5 or +/- 2% of depth; whichever is larger
Maximum elapsed time [sec] 291 73 123 48
Rated ship speed [knot] 6 15 15 10

Since XBT carries no pressure sensor, we need to estimate depth from the elapsed time. The fall-rate equation is as follows.

Z = at + 10E-3 * bt2

Where Z(m) is the depth and t(sec) is the elapsed time.

In addition, coefficients of the fall-rate equation are different by probe types.

Probe Type TSK T-5 TSK T-6 TSK T-7 TSK T-10
Coefficient-a 6.828 6.691 6.691 6.301
Coefficient-b -1.82 -2.25 -2.25 -2.16
* Coefficients listed above are supplied by Sippican, Inc., in USA.

The list of an XBT type used in each cast is as follows.

Cast name Probe Serial No. Probe Type Launcher Converter
BT-009120090702 - T-5 Hand MK-30N
BT-009220090705 - T-5 Hand MK-30N

(1) For sensor's stablility, values of less than 1 m for temperature are replaced by missing values, respectively, based on manufacturer's recommendation.

(2) Quality control
QCed data were added flag according to the NODC (National Oceanographic Data Center) quality control procedure.
1) The gradient check of adjacent depth data
2) The broad range check set up at given ocean space and depth
Please see the site of NODC of the following link for quality control procedure in detail.
In addition, an abnormal value is identified by a visual check, and the data after visual QC is released.
(1) It is reported that T-5 probes produced by Tsurumi Seiki Co. Ltd. (TSK T-5 probes) have a fall-rate bias. Please see the following about publication policy of XBT fall-rate bias correction data.
: YK09-08
: Processed (DMO)-QCed
: JAMSTEC
: Depth, Temperature
:
OCEANS > OCEAN TEMPERATURE > WATER TEMPERATURE

Observation Map

  1. Clicking the icon displays a balloon with observation information.
  2. Then click the observation name, figures will be displayed.
Imagery reproduced from ...

Figures

Data List

File names
BT-009120090702.dat
BT-009220090705.dat
ex_read2.f (Sample Program)
Observation List
The list of observation is shown as follows.
Observation Time and Date Lat. [°] Lon. [°]
BT-009120090702 2009-07-02 22:50 12.8661 143.8466
BT-009220090705 2009-07-05 19:18 11.3684 142.4830
: YK09-08
: Processed (DMO)-QCed
: JAMSTEC

XBT DMO

Each data file contains one line header (meta data) followed by data lines for each cast.
The number of data lines are recorded in the header.
  1. Header part
No. Column Content Format Remarks
1 1 Header ID a1 fixed as '#'
2 3 - 6 Data ID a4 XBT
3 8 - 22 Cruise ID a15
4 33 - 40 Date i8 YYYYMMDD (UTC)
5 42 - 45 Time i4 hhmm (UTC)
6 47 - 55 Latitude i2,a1,f5.2,a1 dd-mm.mmN(S)
7 57 - 66 Longitude i3,a1,f5.2,a1 ddd-mm.mmE(W)
8 68 - 71 Number of data lines i4
9 72 - 73 Terminator - CR+LF
  1. Data part
No. Column Content Unit Format Remarks
1 1 - 11 Depth m f11.1
2 12 - 22 Temperature deg-C f11.2 ITS-90
3 45 - 55 Flag - i11 1 - 7 : space
8 : flag of depth
9 : flag of temperature
10 - 11 : space
* reference : 'Definition of Quality Control Flags'
4 56 - 57 Terminator - - CR+LF
Each contents of the data part is stored in 11 bytes.
Missing value is presented by '-5', and error value is presented by '-9'.

Definition of Quality Control Flags
  1. Depth Flags
  2. 0 - accepted value
    1 - error in recorded depth ( same or less than previous depth )
    2 - density inversion

  3. Observed Level Flags
  4. N - missing value
    0 - accepted value
    1 - range outlier ( outside of broad range check )
    2 - failed inversion check
    3 - failed gradient check
    4 - zero anomaly
    5 - failed combined gradient and inversion checks
    6 - failed range and inversion checks
    7 - failed range and gradient checks
    8 - failed range and zero anomaly checks
    9 - failed range and combined gradient and inversion checks
    A - failed visual check

    * Range and gradient check is performed to XBT data.

QCed data were added flag according to the NODC (National Oceanographic Data Center) quality control procedure, additionally visually checked. Please see the site of NODC of the following link for quality control procedure.

Related Information

Cruise Data Dive Data
: URASHIMA

Update History

2019-09-21
An observation data was registerd.
2017-06-29
An observation data was registerd.
2014-09-11
An observation data was registerd.
2014-03-15
An observation data was registerd.
2012-09-28
An observation data was registerd.